Title: High accuracy optical particle sizing in phase-Doppler anemometry
Authors: Qiu, Hui-He
Jia, W.
Hsu, Chin-Tsau
Sommerfeld, M.
Keywords: Particle size
Measurement-volume effect
Phase-Doppler anemometry
Optical particle sizing
Issue Date: Feb-2000
Citation: Measurement science & technology, v. 11, 2000, p. 142-151
Abstract: A novel structure for phase-Doppler anemometry (PDA) is proposed in order to eliminate the measurement-volume effect (MVE). This new PDA system consists of four receiving detectors in an optimized optical orientation. The method for the determination of the optimized optical orientation angle is introduced. In this method the vector sum of refractive and reflective rays is taken into consideration in describing a dual-mechanism-scattering model caused by a nonuniformly illuminated PDA measurement volume. The constraint of the single-mechanism-scattering model in the conventional PDA is removed. As a result the error caused by MVE, which consists of a Gaussian beam defect and a slit effect, can be eliminated. Analytical and experimental studies were conducted to investigate the accuracy of particle sizing using this new system. Both the experimental results and the simulation based on the generalized Lorenz–Mie theory show that the new method can provide a high-accuracy PDA system free of the MVE.
Rights: Measurement Science and Technology © Copyright (2000) IOP Publishing Ltd. The Journal's web site is located at http://www.iop.org/EJ/journal/MST
URI: http://hdl.handle.net/1783.1/1307
Appears in Collections:MECH Journal/Magazine Articles

Files in This Item:

File Description SizeFormat
e00206.pdf591KbAdobe PDFView/Open

Find published version via OpenURL Link Resolver

All items in this Repository are protected by copyright, with all rights reserved.